| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-55-user had an inaccurate reference: competitor's meter: the end user is comparing results obtained from trividia's bgm system to the results from a competitor's bgm system.Test strip udi# (b)(4).Manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 323 and 219 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 150 mg/dl.Customer was comparing results to another device and stated the results obtained using the truetrack meter were higher; actual meter to meter comparison results were not provided.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer declined to perform a back to back blood test during call on (b)(6) 2018.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 01/30/2021 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :147mg/dl date: (b)(6) 2018 time:05:58am fasting; result 2 :323mg/dl date: (b)(6) 2018 time:10:54pm fasting; result 3 :219mg/dl date: (b)(6) 2018 time:02:33pm fasting; result 4 :173mg/dl date: (b)(6) 2018 time:12:01pm fasting; result 5 :172mg/dl date: (b)(6) 2018 time:06:12am fasting.
|
| |
|
Search Alerts/Recalls
|