BD (SUZHOU) BD INSIGHT-A ¿ CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 382803
Device Problem Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported with the use of the bd insight-a ¿ catheter there was an issue with catheter adopter bent.
 
Manufacturer Narrative
Investigation summary: a device history review was conducted for lot number 8137337.Our records show the reported lot was manufactured in june of 2018 , and determined that this is the only instance of this issue occurring in this batch of insyte-a.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, in response to the damage observed in the sample provided our engineering team launched a review of the manufacturing, inspection, and packaging processes.During their review they were unable to identify any potential abnormalities that could have led to the reported event.Based on these findings, our engineers speculate that the most probable root cause for this occurrence is the application of force to the device during shipment or storage, however this cannot be confirmed to at this time.
 
Event Description
It was reported with the use of the bd insight-a ¿ catheter there was an issue with catheter adopter bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSIGHT-A ¿ CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
MDR Report Key8199714
MDR Text Key131708354
Report Number3006948883-2018-00315
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2021
Device Catalogue Number382803
Device Lot Number8137337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Date Manufacturer Received12/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-