• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR 4.5MM VA-LCP CURVED CONDYLAR PLATE/LEFT; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR 4.5MM VA-LCP CURVED CONDYLAR PLATE/LEFT; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE Back to Search Results
Model Number 03.231.003
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional procodes: hwc, hrs.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A device history record review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during an open reduction and internal fixation (orif) of the right femur on (b)(6) 2018, the arm sleeves of left and right aiming arm were no longer locking securely in the guide.The grooves in the carbon fiber of the aiming arms seem to be wearing out, preventing the metal wings in the sleeves from engaging.The trocar with handle was not properly locking into the sleeves during the sleeve insertion to aiming arm.The case was completed successfully by manually holding the sleeves in the aiming arm during the procedure.There was no surgical delay.There was no patient harm.Concomitant devices: guide (part: unknown, lot: unknown, quantity: 1).This report is for a aiming arm for 4.5mm variable angle locking compression plate (va-lcp) curved condylar plate/left.This is report 1 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: a device history record (dhr) review was conducted: part: 03.231.003, lot: 7549772, manufacturing site: synthes haegendorf, release to warehouse date: 06.May 2011.The device history record shows this lot of 8 pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.H3, h6: a product investigation was conducted.Visual inspection: visual inspection of the returned left aiming arm assembly for va-lcp curved condylar plates revealed no damage to any carbon fiber feature of the device.The connection bolt component was not returned.No damage was noted to the mounting sleeve component¿s threads.Slight damage to the top surface of the mounting sleeve was noted.This damage is consistent with use of the device.No damage was noted the pin or any other feature of the device.The damage noted above would have impacted the ability to assemble with the insertion handle.Functional testing: functional testing could not be performed as the insertion handle was not returned.The received condition did not agree with the complaint description.The complaint could not be replicated during investigation.Dimensional inspection: dimensional inspection of the mounting sleeve¿s hole measured 14.00mm.Measured with ca802.This is within specification of 14.00mm +0.05/-0 per se_370849 rev b.Document/specification review: the relevant drawings were reviewed during investigation: no design issues were noted during investigation.Conclusion: the complaint was not confirmed during investigation.No damage was noted any of the features that interact with the insertion handle, which would have contributed to the complaint condition.No manufacturing issues were identified during investigation.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM FOR 4.5MM VA-LCP CURVED CONDYLAR PLATE/LEFT
Common Device Name
IMPLANT, FIXATION DEVICE, CONDYLAR PLATE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8199724
Report Number2939274-2018-55604
Device Sequence Number1557622
Product Code JDP
UDI-Device Identifier10886982077301
UDI-Public(01)10886982077301
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.231.003
Device Catalogue Number03.231.003
Device Lot Number7549772
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2019
Initial Date Received by Manufacturer 12/04/2018
Supplement Date Received by Manufacturer02/01/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date02/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-