|
The cause for the falsely elevated advia centaur xp alpha-fetoprotein (afp) patient results is unknown.The customer's assay calibration and quality controls (qc) were acceptable.The customer informed siemens that initial afp concentrations were not above the assay measurement range.For example: the sample from (b)(6) had an initial afp result of 7.5 ug/l, when pre-diluted with negative serum pool and tested the next day, the results were 1.7 ug/l (1:10), 14.5 ug/l (1:20), 49.8 (1:200) and eventually yielding a result of > 1000 ug/l.Siemens is investigating the event.The instruction for use (ifu) under the interpretation of results section states the following: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." mdr 1219913-2018-00309 was filed for an elevated afp result on the same patient on a different day.
|
|
Siemens filed mdr 1219913-2018-00308 on 12/27/2018 for a falsely elevated advia centaur xp alpha-fetoprotein (afp) result on a patient sample.On 01/08/2019 - correction b6: there was not an alternate afp test method result for august 31, 2018 as initially stated.Correction d4: the date of expiration for the year in d5 was a typographical error.D5: expiration date - 5/15/2019.On 01/08/2019 - additional information: the customer informed siemens that the patient sample from november, 2018 and tested on the alternate afp test method (4.6 ug/l) had not been tested on the advia centaur xp for afp.On 01/11/2019 - additional information: the advia centaur xp neat results for august 31, 2018, october 07, 2018, and december 06, 2018 were around the assay cutoff range prior to dilutions (actual neat results not provided).The customer was not expecting lower afp results due to the patient's treatment, and the doctor suspected a high dose hook effect as the results before february 2018 rose steeply.The customer diluted the patient samples observing false positive results.The high dilution results did not fit the clinical findings (mrts, pet-ct) or the lower apf results from the other hospital when the patient was transferred.On 01/19/2019 - siemens has concluded the incident investigation.The patient had an increase of afp results in early 2018, and was also undergoing medical treatment.When the results started to decrease, the doctor suspected a high dose hook effect and started to dilute the samples at high dilutions.The end results of the dilution testing showed very high afp results.The patient when transferred to another hospital, and their afp result when tested on an alternate test method was normal.This caused the customer to investigate further, and noted that diluting the samples caused an artificial increase in afp results.In summary, the data suggests the increased and varied afp results were the effect of diluting a sample that was not required to be diluted.The lower advia centaur xp afp from december 12, 2018 neat result of 9.5 ug/l was similar to the alternate afp test method results of 7.5 ug/l and 7.85 ug/l.No product performance issue has been identified by siemens.The instruction for use (ifu) under the interpretation of results section states the following: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the instruction for use (ifu under the dilution section states the following: "for an automatic or manual dilution, the final afp concentration in the diluted sample must be greater than or equal to 15 ng/ml (12.5 iu/ml)." mdr 1219913-2018-00309 was filed for an elevated afp result on the same patient on a different day, and mdr 1219913-2018-00309 supplemental report 1 was filed for corrections and additional information.
|