| Model Number |
174213 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter during a case of excision of mediastinal lesion and thoracoscopy, both device had their coating peeled-off after just one use.The procedure was completed with another device.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Shearing damage of the plastic cable sheath was observed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur where an excessive force or leverage is applied to the device while retracting the clevis assembly and the clevis boot is subsequently damaged.This subsequent damage may result in improper retraction and deployment of the jaws during articulation and may cause shearing of the plastic cable sheath.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|