COVIDIEN LP LLC NORTH HAVEN IDRIVE ULTRA; STAPLE, IMPLANTABLE Back to Search Results
Model Number IDRVULTRA2
Medical Device Problem Codes Misfire (2532); Mechanics Altered (2984)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter during laparotomy at the lower anterior resection of the rectum by laparoscopy, the device cut the tissue but did not staple.The staple line is also incomplete at the distal part.The procedure was converted to an open procedure due to device problem and new staple was used to complete the case.The patient stayed in hospital for 6 days instead of 3-4 days due to the device issue.The procedure was extended by more than 30 minutes due to device issue.The incision was extended by more than one inch.
 
Additional Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.A pmv representative adapter was attached and calibrated correctly.The device was fired on a portable a1 and fired satisfactorily within specification.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDRIVE ULTRA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8199940
Report Number1219930-2018-06881
Device Sequence Number13815157
Product Code GDW
UDI-Device Identifier10884521510579
UDI-Public10884521510579
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K121510
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIDRVULTRA2
Device Catalogue NumberIDRVULTRA2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer05/24/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date06/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-