| Model Number |
IDRVULTRA2 |
| Medical Device Problem Codes |
Misfire (2532); Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter during laparotomy at the lower anterior resection of the rectum by laparoscopy, the device cut the tissue but did not staple.The staple line is also incomplete at the distal part.The procedure was converted to an open procedure due to device problem and new staple was used to complete the case.The patient stayed in hospital for 6 days instead of 3-4 days due to the device issue.The procedure was extended by more than 30 minutes due to device issue.The incision was extended by more than one inch.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary post market vigilance (pmv) led an evaluation of one device.A pmv representative adapter was attached and calibrated correctly.The device was fired on a portable a1 and fired satisfactorily within specification.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|