• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 ASPHERE M SPEC 12/14 36 +8.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 ASPHERE M SPEC 12/14 36 +8.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-53-100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Tissue Damage (2104); No Code Available (3191)
Date of Event 11/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
The patient made an appointment to his doctor after noticing a bump on his thigh.During the surgery, pseudotumors were removed.Doi: 2011.Dor: (b)(6) 2018; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial medwatch , a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
After review of medical records patient was revised to addressed bilateral hip mechanical failure of total hip arthroplasty with pseudotumor formation from metal on metal.There was also some corrosion on the trunnion that was removed using a sponge in saline.Doi: (b)(6) 2011, dor: (b)(6) 2018 left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 12/14 36 +8.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582 0988
MDR Report Key8200050
Report Number1818910-2018-79216
Device Sequence Number11276816
Product Code JDI
UDI-Device Identifier10603295033967
UDI-Public10603295033967
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2016
Device Model Number1365-53-100
Device Catalogue Number136553100
Device Lot Number3276579
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/28/2018
Supplement Date Received by Manufacturer01/01/2019
12/28/2020
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date01/01/2019
01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINNACLE MTL INS NEUT36IDX52OD.; PINNACLE SECTOR II CUP 52MM.; TRI-LOCK BPS SZ 4 HI OFFSET.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient Weight100
-
-