• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC MIXTER ENDOSCOPIC CHOLANGIOGRAPHY SET; GBZ CATHETER, CHOLANGIOGRAPHY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MIXTER ENDOSCOPIC CHOLANGIOGRAPHY SET; GBZ CATHETER, CHOLANGIOGRAPHY Back to Search Results
Catalog Number C-OECS-400-MIXTER
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported foreign matter, piece of hair, was found in the sealed sterile package of a mixter endoscopic cholangiography set during incoming inspection.The device did not make patient contact.
 
Additional Manufacturer Narrative
Corrected data:it was reported a piece of an unidentified particle was found inside of the primary packaging of a mixter endoscopic cholangiography set during incoming inspection.The device did not make patient contact.Device evaluation: the complainant did not return the complaint device to the manufacturer for investigation.However, a set of photos were provided from the customer and show a particle of foreign matter towards the bottom of the package.The package was not opened and prior to use, which suggests the device was out of specification.Investigation: a document-based investigation reviewed the following: complaint history, device history record, documentation, instructions for use, and quality control specifications.A review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.¿how supplied: supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ conclusion: based on the information provided, examination of the device photos, and the results of our investigation, the root cause was determined as cause traced to manufacturing (quality control deficiency).Per the [quality engineering] risk assessment, no further action is required.Appropriate measures were initiated to address this failure mode.As a result, there are new gowning procedures in place that should prevent this failure.The appropriate personnel will be notified and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIXTER ENDOSCOPIC CHOLANGIOGRAPHY SET
Common Device Name
GBZ CATHETER, CHOLANGIOGRAPHY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8200107
Report Number1820334-2018-03708
Device Sequence Number1306755
Product Code GBZ
UDI-Device Identifier00827002080454
UDI-Public(01)00827002080454(17)210227(10)8629195
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2021
Device Catalogue NumberC-OECS-400-MIXTER
Device Lot Number8629195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Initial Date Received by Manufacturer 12/05/2018
Supplement Date Received by Manufacturer02/06/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date03/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-