• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS NAVARRE UNIVERSAL DRAINAGE CATHETERS WITH NITINOL; CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS NAVARRE UNIVERSAL DRAINAGE CATHETERS WITH NITINOL; CATHETER, IRRIGATION Back to Search Results
Catalog Number NNU6LP15
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was not provided, therefore, the device history records were not reviewed.The return of the device is pending.The investigation is currently underway.
 
Event or Problem Description
It was reported that the device was allegedly cracked.Reportedly, the device was removed and replaced.There was no reported patient injury.
 
Event or Problem Description
It was reported that the device was allegedly cracked.Reportedly, the device was removed and replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one navarre locking pigtail 6 fr x 15cm universal drainage catheter was returned for evaluation.Functional, gross visual, microscopic visual and tactile evaluations were performed.The investigation is unconfirmed for crack in device, as no cracks on the device were observed during sample evaluations.The definitive root cause could not be determined based upon available information, as the alleged device issue was unconfirmed.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE UNIVERSAL DRAINAGE CATHETERS WITH NITINOL
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8200167
Report Number3006260740-2018-03748
Device Sequence Number13973425
Product Code FGE
UDI-Device Identifier00801741036965
UDI-Public(01)00801741036965
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K955110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNNU6LP15
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer02/13/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date03/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-