| Model Number |
42544 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to the procedure, the physician found there was a ¿pressure problem.¿ the valve was to be replaced and there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent and met the requirements for reflux and leak testing.However, the valve did not meet the requirements for pressure-flow and preimplantation testing.There was proteinaceous debris on the interior of the valve.The instructions for use cautions, ¿care must be taken to ensure that p articulate contaminants are not introduced into shunt components during preimplantation testing or handling.Introduction of contaminants could result in improper performance (overdrainage or underdrainage) of the shunt system.Particulate matter, which enters the shunt system, may also hold pressure/flow controlling mechanisms open, resulting in overdrainage.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that the physician tested the shunt pressure before implantation, and they couldn't get a normal pressure.
|
| |
|