| Model Number |
42866 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, the valve broke and leaked.The valve was never implanted and there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient weight provided.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the cause of the damage was not clear.It was stated the valve broke during the patency test prior to the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for valve flux, siphon, reflux, pressure flow and pre-implantation testing.However, the valve did not meet the requirement for leak testing.The valve failed leak testing due to a tear on the casing of the delta chamber.It is unknown how or when this damage occurred.The instructions for use cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity¿¿ debris was observed on the interior and exterior of the valve.The instructions for use cautions, ¿care must be taken to ensure that particulate contaminants are not introduced into shunt components during preimplantation testing or handling.Introduction of contaminants could result in improper performance of the shunt system.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|