Model Number METRIX AIR
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Replaced test strips only due to improper storage.Send control solution also.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 275, 269 and 115 mg/dl.Customer also stated she had tested that morning and obtained a result of 200 mg/dl fasting.Customer stated the high readings started that tuesday.The customer's expected fasting blood glucose test result range is 70 - 98 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 02/24/2020 and customer stated the vial was opened recently.The meter memory was reviewed for previous test result history: result 1:275mg/dl 9:50am (b)(6) fasting; result 2:269mg/dl 935am (b)(6) fasting; result 3:115mg/dl 10:11am (b)(6) fasting; result 4:88mg/dl 9:16am (b)(6) fasting; result 5:90mg/dl 10:30am (b)(6) fasting.Customer state that meter are reading higher than usual.Customer does not know what her a1c is customer stated she had not had an a1c test.Customer tested this morning and got a reading of 200mg/dl fasting.High readings started this tuesday.Customer has not had any changes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8200591
Report Number1000113657-2018-01292
Device Sequence Number13848612
Product Code NBW
UDI-Device Identifier00311917173894
UDI-Public(01)00311917173894
Combination Product (Y/N)N
Initial Reporter StateOH
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/24/2020
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2906S
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2018
Initial Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-