| Catalog Number |
103579 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during priming of a polyflux 170h dialyzer, black particles were observed inside the filter ¿on the blood side of the dialyzer¿.No particles were observed in the set.The set and filter were changed prior to starting treatment.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.During visual inspection by naked eye, the product was found to be wet and small black particulate matter was observed in the header on one side of the sample.The reported issue could be confirmed.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|