ABBOTT POINT OF CARE I-STAT EC8+ CARTRIDGE Back to Search Results
Catalog Number 600-9007-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Anemia (1706); Renal Failure (2041)
Date of Event 12/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat ec8+ cartridges that yielded suspected discrepant results on a (b)(6) male patient with history of anemia, secondary to chronic renal failure.There was no patient information available at the time of this report.Customer reports an i-stat value of 15% pcv.An initial manual pcv result was 36%, a second manual pcv results was 45%.Dates and times of the test were not documented.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.It is suspected that the manual pcv determined may be a hemoglobin result.However, there is not enough information to support the cause.Nor it is clear at this time what "manual method" was being used.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 02/06/2019.A review of the device history record confirmed the lot passed finished goods release criteria.Retain cartridge test results met the acceptance criteria found in q04.01.003 rev.Ac, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EC8+ CARTRIDGE
Common Device Name
EC8+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8200845
Report Number2245578-2018-00321
Device Sequence Number137921
Product Code CHL
UDI-Device Identifier10054749000071
UDI-Public10054749000071
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K940918
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/13/2019
Device Catalogue Number600-9007-25
Device Lot NumberK18165
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer02/06/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date02/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age13 YR
-
-