BARD PERIPHERAL VASCULAR, INC. ULTRASCORE FOCUSED FORCE PTA BALLOON Back to Search Results
Model Number US35130520
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 10/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No medical images have been made available to the manufacturer.Medical records have been provided and a review is currently underway.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 03/2020), (manufacturing date: 04/2018).
 
Event or Problem Description
It was reported through the results of a clinical trial that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter, vessel dissection was identified after one inflation in the right superficial femoral artery.Reportedly, the dissection was treated with a stent.The event was considered to be resolved.The patient was discharged the same day as the index procedure.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was not returned, but medical records were provided.Based on the medical records, dissection following the use of a balloon can be confirmed.However, the investigation is inconclusive, as no objective evidence has been provided to confirm an alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 03/2020), (manufacturing date: 04/2018).
 
Event or Problem Description
It was reported through the results of a clinical trial that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter, vessel dissection was identified after one inflation in the right superficial femoral artery.Reportedly, the dissection was treated with a stent.The event was considered to be resolved.The patient was discharged the same day as the index procedure.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE FOCUSED FORCE PTA BALLOON
Common Device Name
FOCUSED FORCE PTA BALLOON
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8200874
Report Number2020394-2018-02255
Device Sequence Number2056509
Product Code PNO
UDI-Device Identifier00801741116803
UDI-Public(01)00801741116803
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K172571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUS35130520
Device Catalogue NumberUS35130520
Device Lot NumberGFCN4129
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/30/2018
Supplement Date Received by Manufacturer01/18/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date02/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient Weight118
-
-