|
|
| Catalog Number |
72201693 |
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an shoulder arthroscopy, the patient suffered cartilage damages in the humerus as a consequence of an existing gap between the 5.00 mm cannula tip-end (which has a sharp edge) and the arthroscope when connected.As this adverse consequence was noticed after the procedure, the surgery had been finished with the same device, in a timely manner.It is still unknown: (i) the outcome and evolution of the patient; and (ii) if there was any necessary revision surgical intervention to preclude any permanent damage to the cartilage.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to no product return the complaint could not be ultimately confirmed.Definitive conclusions, concise investigation and evaluation are not possible without product to evaluate.No further actions pursued at this time.If objective evidence, relevant information, package or product becomes available to assist with evaluation, the complaint will certainly be revisited.
|
| |
|
Search Alerts/Recalls
|
|
|