Model Number 5F061503CS
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 10/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Medical records review: the patient with history of dry gangrene and total occlusion of the right superficial femoral artery was scheduled for vascular intervention.Two stents were placed and there was no reported complication.Approximately one week post hospital admission, the patient was right critical limb ischemia required amputation.The right common femoral artery was accessed and an arteriography demonstrated 4cm occlusion of the left superficial femoral artery.The decision was made to place a stent in the left superficial femoral artery prior to right below the knee amputation.Hemostasis was achieved and the patient was transported in stable condition.(02/2018).
 
Event or Problem Description
It was reported through the results of a clinical trial, that approximately one month post angioplasty procedure in the distal superficial femoral artery (sfa) with a focused force percutaneous transluminal angioplasty (pta) balloon, the patient had two vascular stents placed in the right sfa for total occlusion.Approximately one week post vascular stent placement, the patient underwent right below the knee amputation due to dry gangrene and critical limb ischemia.The vascular stents were stenosed at the time of the amputation.The current patient status was not provided.
 
Event or Problem Description
It was reported through the results of a clinical trial, that approximately one month post angioplasty procedure in the distal superficial femoral artery (sfa) with a focused force percutaneous transluminal angioplasty (pta) balloon, the patient had two vascular stents placed in the right sfa for total occlusion.Approximately one week post vascular stent placement, the patient underwent right below the knee amputation due to dry gangrene and critical limb ischemia.The vascular stents were stenosed at the time of the amputation.The current patient status was not provided.
 
Additional Manufacturer Narrative
Manufacturing review: the lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.Investigation summary: as no x-ray images were provided the reported in stent restenosis could not be verified.An alleged device relation could not be determined.Based on the information available the investigation will be completed with inconclusive result.In this case the lesion was pre and post dilated, and the stent was successfully placed without experience of malfunction or failure.Based on the information available a definite root cause for the event reported could not be identified.Labeling review: in reviewing the current version of the labeling supplied with this product it was found that the ifu sufficiently addresses the potential risk.The ifu states that 'arterial occlusion/restenosis of the treated vessel' as well as 'amputation' are potential adverse events that may occur.Based on the ifu pre-dilatation of the lesion as well post stent expansion with a balloon dilatation catheter is recommended.
 
Event or Problem Description
It was reported through the results of a clinical trial, that approximately one month post angioplasty procedure in the distal superficial femoral artery (sfa) with a focused force percutaneous transluminal angioplasty (pta) balloon, the patient had two vascular stents placed in the right sfa for total occlusion.Approximately one week post vascular stent placement, the patient underwent right below the knee amputation due to dry gangrene and critical limb ischemia.The vascular stents were stenosed at the time of the amputation.The current patient status was not provided.
 
Additional Manufacturer Narrative
Manufacturing review: the lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.Investigation summary: as no x-ray images were provided the reported in stent restenosis could not be verified.An alleged device relation could not be determined.Based on the information available the investigation will be completed with inconclusive result.In this case the lesion was pre and post dilated, and the stent was successfully placed without experience of malfunction or failure.Based on the information available a definite root cause for the event reported could not be identified.Labeling review: in reviewing the current version of the labeling supplied with this product it was found that the ifu sufficiently addresses the potential risk.The ifu states that 'arterial occlusion/restenosis of the treated vessel' as well as 'amputation' are potential adverse events that may occur.Based on the ifu pre-dilatation of the lesion as well post stent expansion with a balloon dilatation catheter is recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT 5F VASCULAR STENT SYSTEM
Common Device Name
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe AZ 76227
GM  76227
MDR Report Key8201147
Report Number9681442-2018-00224
Device Sequence Number16178389
Product Code NIP
UDI-Device Identifier00801741120206
UDI-Public(01)00801741120206
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P070014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,study
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2020
Device Model Number5F061503CS
Device Catalogue Number5F061503CS
Device Lot NumberANCN1563
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/30/2018
Supplement Date Received by Manufacturer01/18/2019
05/21/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date01/30/2019
05/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Disability;
Patient Age73 YR
Patient Weight58
-
-