Model Number FZ120-200T-25
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vilex received call on (b)(6) 2018, regarding a broken 12mm fuze.(b)(6) 2015 - 12mm fuze (fz120-200t-25) was implanted in patient.(b)(6) 2018, surgery was performed to have the broken implant removed and a fusion revision was completed.Based on x-rays and discussion with the doctor, this was a non-union and the patient walked on it until it broke.A review of complaints was completed and none were found on any size of the 12mm fuze.A review of non-conformances was completed and none were found on this item, fz120-200t-25, lot 7018.If any additional information becomes available, vilex will file a supplemental report.
 
Event or Problem Description
Fuze nail broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUZE INTRAMEDUALLARY NAIL
Common Device Name
FUZE
Manufacturer (Section D)
VILEX, INC.
111 moffitt street
mcminnville 37110
MDR Report Key8201210
Report Number1051526-2018-00007
Device Sequence Number10814819
Product Code HSB
UDI-Device Identifier00841731107275
UDI-Public(01)00841731107275(10)7018
Combination Product (Y/N)N
PMA/510(K) Number
K102413
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFZ120-200T-25
Device Catalogue NumberFZ120-200T-25
Device Lot Number7018
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2018
Initial Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-