Model Number 0094160
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the eyelet on the catheter was cut closer on one side than on the other side.
 
Event or Problem Description
It was reported that the eyelet on the catheter was cut closer on one side than on the other side.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.Ten samples were received open and without original packaging.Visual observation noted no defects, burrs, or sharp lumps.Eye location relative to the lateral plane was measured.The shaft wall height was expected to be equal on either side of the eye.The catheter shaft walls heights ranged from 0.107 inches to 0.172 inches.The maximum observed discrepancy in shaft wall height averaged over the two shaft wall heights is 0.0013 inches.The eye location relative to the shaft diameter does not affect function of the device and is unspecified.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® ALL PURPOSE URETHRAL CATHETER
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8201439
Report Number1018233-2018-06226
Device Sequence Number223188
Product Code KOD
UDI-Device Identifier00801741034039
UDI-Public(01)00801741034039
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number0094160
Device Catalogue Number0094160
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2018
Initial Date Received by Manufacturer 12/06/2018
Supplement Date Received by Manufacturer04/18/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date05/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-