OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", 24 FR; ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL
|
Back to Search Results |
|
| Model Number |
WA22703S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not returned to olympus for evaluation.The cause of the reported complaint cannot be confirmed.As a preventive measure, the device instructions for use warns against bending the distal tip.The instructions for use also cautions, ¿if signs of wear become visible during the procedure, replace the product.¿ the instructions for use also have pre-procedure directions for inspecting both the device and its packaging for damage and deformation.
|
| |
|
Event or Problem Description
|
|
Olympus was informed via medwatch report (b)(4) that during a cystoscopy with transurethral resection of prostate, the loop of the device broke in the middle and a fragment fell into the bladder.The fragment was washed out of the patient by flushing the bladder.There was no reported patient injury.
|
| |
|