| Model Number |
9734228 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The driver was returned to the manufacturer for analysis.Analysis found that the post on one of the support arms of the driver was missing.Otherwise, with the remaining markers attached and fully seated, the driver returned good geometry and divot error readings.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside a procedure.It was reported that the gray navlock driver was damaged.There was no patient present when this issue was identified.
|
| |
|