WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM LCP PROXIMAL TIBIA PLATE 14 HOLES/262MM-RIGHT; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
240.046 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional patient identifier: (b)(6).Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, the patient underwent hardware removal due to infection.During procedure, the plates had to be cut away to remove the screws.All implants were removed successfully.However, there was difficulty removing the fragments that were generated from the broken devices and trephine was used to remove the screw head.No other device was use before using the trephine.Original implantation was on (b)(6) 2015, for an open reduction and internal fixation (orif) of the proximal tibia.The procedure was completed successfully.There was no surgical delay.The patient outcome was stable.This report addresses the intra-op event of the difficulty removing the screw, while (b)(4) will capture post op event that involves the infection/one broken screw.This report is for one (1) 4.5mm lcp proximal tibia plate.This is report 2 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history records.Part mfg date: 26-dec-2012.Part exp.Date: n/a.Manufacturing location: depuy synthes: elmira.Lot number: 7138154.Part number: 240.046.Dhr record review: a review of the device history record (dhr) revealed no non-conformances.The dhr shows lot#: 7138154 of 4.5mm lcp proximal tibia plate 14 holes/262mm-right was processed through the normal manufacturing and inspection operations with no non-conformances or rework noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Device history batch null, device history review: this lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during manufacture that would contribute to this complaint condition.
|
| |
|