Model Number 16-02-80
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2018
Event Type  malfunction  
Manufacturer Narrative
The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective stirrer motor.The technician replaced the motor to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The defective stirrer motor was returned for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message on the patient side during procedure.There was no report of patient injury.
 
Event Description
See initial.
 
Manufacturer Narrative
The affected parts has been requested back to the manufacturer site for a detailed investigation.Results revealed that the stirrer motor had defective electronics.Furthermore, in accordance with the replacement conducted on site, investigation results confirmed also the involvement of the distributor board.The protective resistors r1 and r2 were high-impedance and two sockets were found to be defective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8202254
MDR Text Key131912367
Report Number9611109-2018-01445
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received05/02/2019
Supplement Dates FDA Received05/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-