| Model Number |
31317062U |
| Medical Device Problem Codes |
Material Separation (1562); Vibration (1674)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported via a manufacturer representative that the irrigation channel snapped during a resection of acoustic neuroma procedure.The irrigation channel could not be used with the bur guard.The surgeon decided to use another handpiece with irrigation.The surgeon stated that the bur felt like it was vibrating/wobbling and then it snapped.The distal part of the bur was retrieved.The proximal part of the bur was stuck inside the drill.There was a delay in the procedure as a result of this event.There was no patient impact or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis found that the shaft of the bur broken.The measure from bur tip to break was 1.375".The break configuration is consistent with a shear or bend force.If information is provided in the future, a supplemental report will be issued.
|
| |
|