MAUDE Adverse Event Report: BAXTER HEALTHCARE - CARTAGO CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CARTAGO CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2H8401
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that when the "slide or roller" clamp on a clearlink solution set was closed, it sliced the tubing and caused a leak.This was identified while in use on a patient to deliver an unspecified chemotherapeutic drug with an infusion pump.The reporter stated that there was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Correction : (initial reporter first name corrected, previously reported as jun) additional information : the actual sample was received and evaluated.The sample was contaminated with chemotherapy.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.No functional testing was performed due to the status of the returned sample.The reported problem was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
MDR Report Key8202514
Report Number1416980-2018-08050
Device Sequence Number1656731
Product Code FPA
UDI-Device Identifier00085412049090
UDI-Public(01)00085412049090
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K112893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/26/2023
Device Catalogue Number2H8401
Device Lot NumberR18F26051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2018
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer01/25/2019
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CHEMOTHERAPY DRUG; FLOGARD PUMP
-
-