Model Number 9733856R
Medical Device Problem Codes Mechanical Problem (1384); Output Problem (3005)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the system was running slow due to backed up exams.The old exams were deleted from the archive.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system.It was reported the system was moving very slow and then became unresponsive on the blue screen.A soft reboot was completed, but the system was still slow.It was noted the patient screen was showing over 70% full.Technical services (ts) recommended deleting patients and checking/deleting old versions of software.This issue was noted outside of a procedure, and there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key8202906
Report Number1723170-2018-06490
Device Sequence Number1710460
Product Code HAW
UDI-Device Identifier00613994603678
UDI-Public00613994603678
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number9733856R
Device Catalogue Number9733856R
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2018
Initial Report FDA Received Date12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-