| Model Number |
1883670HS |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported via a manufacturer representative that the bur was broken during the procedure.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the device found that there was a residue consistent with biological contaminants on the device.Visually, the spiral wrap was broken.The spiral wrap was twisted in on itself in a clockwise direction.The portion that became detached measured 2.04¿ long.The distal end of the outer tube inside diameter was worn on one side with corresponding damage to the inner assembly.If information is provided in the future, a supplemental report will be issued.
|
| |
|