| Catalog Number |
482802745 |
| Health Effect - Clinical Codes |
Injury (2348); No Code Available (3191)
|
| Date of Event |
12/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the screw head came off post-operatively.Per the correspondence with rep, the patient underwent revision surgery on (b)(6) 2018 to replace hardware of bigger size.This will be reported as a serious injury due to medical intervention.The rod was also replaced in revision surgery, however, no issue was reported for the rod.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: x-ray image review, visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment results: the reported event of was confirmed via x-ray images and visual inspection of the returned device.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Inspection of the returned device confirmed deformation on the tulip head locking ring and on the head of the screw shank.Deformation of tulip head locking ring indicates that the screw shaft was pressing against deformed side of the tulip head likely due to over angulation on screw.Deformation on the screw shank and x-ray images also indicate over-angulation on the screw.Conclusion: the likely root cause of the reported event is the over-angulation of the screw.Other plausible cause could be poor bone quality of the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the screw head came off post-operatively.Per the correspondence with rep, the patient underwent revision surgery on (b)(6) 2018 to replace hardware of bigger size.
|
| |
|