Model Number 030450
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic hepatectomy procedure, when using on a blood vessel, the first firing stopped midway.They replaced the reload and continued using the handle to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the loading was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the 2 cm cut line.Microscopic evaluation noted that the loading unit had damage to the cutting edge of the knife blade.Functionally the loading unit was loaded into a post market vigilance instrument, the interlock was over ridden, and the loading unit was applied to test media.All remaining staples were placed, and test media was cleanly transected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damaged knife blade may occur when an obstacle has been incorporated in the jaws during application.The information booklet which accompanies each product shipment cautions the user; ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8203333
Report Number1219930-2018-06920
Device Sequence Number18890572
Product Code GDW
UDI-Device Identifier20884523002819
UDI-Public20884523002819
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number030450
Device Catalogue Number030450
Device Lot NumberN7B0589KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Initial Date Received by Manufacturer 12/04/2018
Supplement Date Received by Manufacturer05/07/2019
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date05/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-