• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Catalog Number 284002
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the device was received at the service center and evaluated.The defect reported by the customer has been verified and repaired.This complaint is confirmed.Salt was found inside the pump; the unit was cleaned.The cpu board was found to be defective and was replaced.The repaired device passed all tests and is fully functional.The root cause is saline fluid getting into the pump which resulted in the cpu board failure.A review of the device history record indicated that this lot of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point in time, no corrective action is required; and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
 
Event Description
It was reported by the affiliate via mail that pre-operatively the device had an unknown problem.They used another device to complete the procedure and there was no known patent consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8203522
MDR Text Key131709085
Report Number1221934-2018-55865
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-