| Catalog Number |
48328500 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tip of screwdriver broke off intra-op.All broken fragments were retrieved successfully.No delay, no medical intervention, and no adverse consequences were reported with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment result: the reported event was confirmed via visual inspection of the returned device.Per visual inspection, wear and deformation was identified around broken tip.The device was approximately 8 years old at the time of the event.Manufacturing files were reviewed and no revealed anomalies were found.Conclusion: the most likely root cause of the tip fracture is normal wear due to device age.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of screwdriver broke off intra-op.All broken fragments were retrieved successfully.No delay, no medical intervention, and no adverse consequences were reported with this event.
|
| |
|