MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5532-G-513
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Liner would not lock.Update (b)(6) 2018: reporter indicated "yes" to surgical delay.
 
Additional Manufacturer Narrative
Reported event: an event regarding seating/locking issues involving a triathlon insert was reported.The event was confirmed through evaluation of the returned device.Method & results: -device evaluation and results: a review was completed by the materials engineer who concluded: damage observed on the insert consistent with attempted implantation and contact against a hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.-medical records received and evaluation: a medical review was not performed because no medical information was provided.-device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.-complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: the insert component was returned for review - a review was completed by the materials engineer who concluded: damage observed on the insert consistent with attempted implantation and contact against a hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.While it can be confirmed that a seating issue did take place through return of the component, the root cause cannot be determined as insufficient information was provided.Further information such as operative reports/images of the device taken when the alleged event was identified is required to complete the investigation.
 
Event or Problem Description
Liner would not lock.Update 05/december/2018: reporter indicated "yes" to surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8203751
Report Number0002249697-2018-04158
Device Sequence Number1670390
Product Code MBH
UDI-Device Identifier07613327050790
UDI-Public07613327050790
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number5532-G-513
Device Lot Number88602018D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer01/11/2019
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date02/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age52 YR
Patient Weight92
-
-