• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD SYRINGE WITH HYPODERMIC NEEDLE PRECISIONGLIDE¿; CONVENTIONAL SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD SYRINGE WITH HYPODERMIC NEEDLE PRECISIONGLIDE¿; CONVENTIONAL SYRINGE Back to Search Results
Catalog Number 309643
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event or Problem Description
It was reported that during use of the bd syringe with hypodermic needle precisionglide¿ there was an issue with foreign matter on the top of the syringe.
 
Event or Problem Description
It was reported that during use of the bd syringe with hypodermic needle precisionglide¿ there was an issue with foreign matter on the top of the syringe.
 
Additional Manufacturer Narrative
H.6.Investigation summary: three photos and one loose syringe were received and visually evaluated.Based on the photos, the syringe was manipulated prior to arriving for evaluation.Photos depict unidentified liquid residue inside the barrel in the fluid path and on the stopper along with foreign matter particulate.The sample received had its stopper in the bottom out position.Upon inspection of the syringe, foreign matter was observed in the fluid path between the stopper and the bottom of the barrel.The foreign matter was a large brown particle, larger than level 2 in size.The foreign matter was wet with unidentified liquid residue inside the barrel.No further evaluation could be performed due to potential risk associated with handling the sample.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.The potential root cause for the foreign matter defect cannot be determined based on photos and the sample provided.The sample was received manipulated, not in its original packaging.Therefore, based on the available information it is not possible to conclude the defect was caused by a manufacturing process.The defect could not be confirmed to have originated at the manufacturing plant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYRINGE WITH HYPODERMIC NEEDLE PRECISIONGLIDE¿
Common Device Name
CONVENTIONAL SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key8203788
Report Number1213809-2018-00935
Device Sequence Number218235
Product Code FMF
UDI-Device Identifier30382903096436
UDI-Public30382903096436
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2022
Device Catalogue Number309643
Device Lot Number7335909
Initial Date Received by Manufacturer 12/10/2018
Supplement Date Received by Manufacturer12/10/2018
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-