| Catalog Number |
1012530-40 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection was performed on the returned device which identified the foreign material is part of the tyvek packaging material.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A query of the complaint database for the reported lot was performed and revealed no other incidents.The investigation determined the loose material in the sealed header bag appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular (av) will continue to trend the performance of these devices.
|
| |
|
Event or Problem Description
|
|
It was reported that a white foreign material was noted inside the package of an 8.0 x 40 mm absolute pro vascular stent delivery system (sds).There was no damage noted in the inner pouch or the device itself.There was no patient involvement.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|