Catalog Number 542-11-50E
Medical Device Problem Codes Malposition of Device (2616); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that tha was done with trident ha shell and x3 liner and stem.The shell was dislocated in about one year after tha.Therefore, all devices were removed and revision surgery was done.The removed polyethylene liner was yellowish discolored.Update dec 17, 2018- cause of revision shell was dislocated (malposition).
 
Additional Manufacturer Narrative
An event regarding loosening involving a trident shell was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection of the returned device noted the following: the shell was returned with the liner locked into it and with the fixation screw protruding through the screw hole.Ha coating is still present on the outer surface and there is some areas with slight bony ingrowth.Apart from this, the shell and screw appear unremarkable.-medical records received and evaluation: not performed as no medical information was received for review product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that tha was done with trident ha shell and x3 liner and stem.The shell was dislocated in about one year after tha.Therefore, all devices were removed and revision surgery was done.The removed polyethylene liner was yellowish discolored.Update dec 17, 2018- cause of revision shell was dislocated (malposition).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL HA CLUSTER 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8203902
Report Number0002249697-2018-04161
Device Sequence Number1630823
Product Code LPH
UDI-Device Identifier07613327037265
UDI-Public07613327037265
Combination Product (Y/N)N
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Catalogue Number542-11-50E
Device Lot Number61473503
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer02/20/2019
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date03/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
Patient Weight40
-
-