Model Number ZXT300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Halo (2227); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: (b)(6) 2018.(b)(4).Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the review.The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that tecnis symfony toric multifocal intraocular lens (model zxt300 +28.5 diopter) was explanted from patient¿s right eye in secondary procedure because of patient experiencing halos.Patient ended up having monofocal toric.Additional information was received, and it was learnt that patient was experiencing high hyperopia, astigmatism, narrow angle, significant haloes and glare with lights.Pre op refraction od:+6.0-2.75x35.Primary surgery was uncomplicated.There was no patient injury.Patient was doing well.No further information provided.
 
Additional Manufacturer Narrative
The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8203980
Report Number9614546-2018-01209
Device Sequence Number13518976
Product Code POE
UDI-Device Identifier05050474606555
UDI-Public(01)05050474606555(17)210819
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/19/2021
Device Model NumberZXT300
Device Catalogue NumberZXT300U285
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 12/07/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-