| Model Number |
945ECLC |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care professional (hcp) reported that the controller was having issues with acquiring motor evoked potentials (mep).No response was achieved after having a voltage up until 700 volts.The patient would sometimes react to the stimulus and sometimes there is no reaction.No back-up device was used and there were no issues during the procedure but mep was not run at the time.There was no known patient impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis found no fault with the device.The complaint could not be duplicated.The device was tested and passed manufacturing s pecifications.The complaint could not still be duplicated in the burn in.(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information from a health care professional (hcp) reported that the event occurred during the procedure.There was no delay in the procedure as the same device was used.There was no patient impact.
|
| |
|