Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Date of Event 10/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device has not been returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zlb00 23.5 diopter intraocular lens (iol) was explanted for the patient's eye due to blurry vision caused by unaddressed cylinder.Reportedly, the calculations were off.The lens was replaced with a model zct225 25.0 diopter lens.There was no patient harm and no other surgical intervention other than the exchange.
 
Additional Manufacturer Narrative
Additional information: the customer provided further information on the event.Explanting doctor was dr.(b)(6).Patient is a female, date of event was (b)(6) 2018, a specific date not provided.Lens was implanted in the patient's right eye.Patient had astigmatism pre op.Confirmation that a toric lens was used as a replacement, a model zct225, 25.0 diopter.Patient did very well post op.The following sections have been updated accordingly: gender/sex: female.(b)(6).Date of event: (b)(6) 2018; an exact date was not provided/known.Entered as (b)(6) 2018.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8204064
Report Number9614546-2018-01211
Device Sequence Number1592923
Product Code MFK
UDI-Device Identifier05050474560741
UDI-Public(01)05050474560741(17)200329
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/06/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/29/2020
Device Model NumberZLB00
Device Catalogue NumberZLB00U0235
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/28/2018
Supplement Date Received by Manufacturer01/04/2019
10/25/2020
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date01/21/2019
11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Date Report Sent to FDA01/01/2005
-
-