| Model Number |
FT3000 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the procedure, the doctor was using the device and a flame came out of the pencil tip.They changed the tip and it happened again.They opened a new device and it worked fine.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned products met specification as received.Visual inspection of the electrode found the coating to have partially worn away and the tip of the e1450x electrode to be charred.The ft3000 did not have any visual defects.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.Visual inspection found part of the coating to have worn away and the electrode to be charred.The purpose of the coating is to limit the tissue from sticking to the electrode.The coating erodes as the electrode is activated.The tip of the electrode also discolors when activated and used on tissue.This effect is a normal part of electrosurgery and is not considered a defective product.The electrode showed no sign of heat damage.The electrode was attached to a pencil and activated at 60w.No flame or abnormal effects "was" noted.The investigation found the device to function normally.The ifu warns: the sparking and heating associated with electrosurgery can provide an ignition source.Observe fire precautions at all times: when using electrosurgery in the same room with gases or flammable substances, prevent pooling of fluids and the accumulation of gases under surgical drapes or near the surgical site.Tissue buildup (eschar) on the tip of an active electrode poses a fire hazard, especially in oxygen-enriched environments, such as in throat or mouth procedures.Eschar plus high oxygen may create embers.Keep the electrode clean and free of all debris.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|