MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 59410882
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2018
Event Type  malfunction  
Event Description
It was reported that when the customer removed the carrier#2, the film dressing adhered to the carrier#2 and got peeled off from the frame.
 
Manufacturer Narrative
Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The new information states that the dressing was never used by the patient, therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Manufacturer Narrative
We have now completed our investigation into this complaint.The returned samples have been assessed and the reported issue can be confirmed.In april 2017 the manufacturing line where iv3000 window frame is produced was part of a continuous improvement project.This project included some modifications to the manufacturing line to improve the consistency and presentation of both the adhesive film and the window frame carrier paper.At the same time improvements were also made to the production line that produces the raw material used to manufacture the final product.This work included an upgrade of the sensors used to control the tension within the bulk rolls.As well as this a reduction in the overall length of the rolls was implemented as this also helped to reduce the faults occurring in the final product a combination of the upgrade and modification to the manufacturing lines has shown measurable improvements in the consistency and quality of the product.Smith and nephew are continually investigating ways to develop and improve our products and we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 WINDOW FRAME 10X12CM CTN 50
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key8204252
MDR Text Key131706832
Report Number8043484-2018-00465
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Type of Report Initial,Followup,Followup
Report Date 06/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number59410882
Device Lot Number1732
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/11/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received12/11/2018
12/11/2018
Supplement Dates FDA Received02/13/2019
06/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-