| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional procode = dro.The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty transistor on the processor/bridge/pace board.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self test for defib and pace functions.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|