Follow-up #1 and final report submitted to update sections d.3, g.1, g.4, g.7, h.2, h.3, h.6, h.10 and h.11 based on the results of investigation.Reported event: it was reported that during reaming trigger on mics hand piece was getting stuck.Product evaluation and results: product inspection could not be performed as the product was not available for evaluation.Product history review: device history records indicate (b)(4) devices were manufactured under lot k06ny and accepted into final stock on 3/22/2016.No non-conformances were identified during inspection.Complaint history review: a review of complaints related to p/n 209063, prodex lot k06ny shows no additional complaint(s) related to the failure in this investigation.Conclusions: the failure mode could not be confirmed because the part was not available for evaluation.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated that there have been no nc and capa are associated with the failure mode reported in this event.
|