• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION CR OXIN FEM SZ5 RT; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLY}METAL}POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION CR OXIN FEM SZ5 RT; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLY}METAL}POLY Back to Search Results
Catalog Number 71421225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 07/25/2014
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient started with erythema swelling suprapatellar area for 2 weeks.She was complained of pain due to cellulitis.The patient had pain stiffness and failed post attempts and manipulation of the right knee.She was admitted for iv vancomycin administration for suspected deep joint prosthetic infection.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned for evaluation.A review of the complaint history for the listed parts revealed no prior complaints for the listed batches.A review of the manufacturing or product specifications did not reveal any abnormalities that could have contributed to the reported issue.Product was sterilized according to sterilization release documentation from quality control.Our clinical/medical team concluded:although it is unknown if the patient¿s immune disorder and comorbidities contributed to the abundant scar tissue/arthrofibrosis for which the mua was performed, arthrofibrosis is a known occurrence and possible complication post tka and not a failure of the prosthesis.There was insufficient supporting documentation of the reported cellulitis and antibiotic therapy; however, based on the lab results the reported erythema and suprapatellar swelling could have been secondary to the mua or other factors but not likely a result of the well-fixed prosthesis or a deep joint infection.The root cause of the gross instability secondary to medial-lateral ligament laxity for which the right tka revision was required cannot be concluded; however, the loss to follow up for approximately 2 years likely contributed to worsening symptoms and a prosthetic malfunction was not supported.The patient impact beyond the reported interventions, revision surgery, and expected post-procedural rehab cannot be concluded.No further medical assessment can be rendered at this time.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION CR OXIN FEM SZ5 RT
Common Device Name
PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLY}METAL}POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8205232
Report Number1020279-2018-02948
Device Sequence Number1887409
Product Code JWH
UDI-Device Identifier00885556030424
UDI-Public00885556030424
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K043440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number71421225
Device Lot Number13MM00338
Initial Date Received by Manufacturer 11/28/2018
Supplement Date Received by Manufacturer11/28/2018
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date06/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age55 YR
-
-