JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); No Code Available (3191)
Date of Event 09/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If explanted; give date: n/a (not applicable).Lens remains implanted.(b)(4).Device evaluation: product testing could not be performed because the lens remains implanted.The complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search of complaints revealed no additional investigation requests have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.Attempts have been made to obtain additional information; however, to date a response has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the patient has bilateral lens implants.An intraocular lens, (iol) model zxt225 15.5 diopter was implanted in the patient's right eye (b)(6) 2018.Patient is not happy with the results and has requested an exchange of the lens.Reportedly, she has been experiencing spider-web vision since surgery.She stated that she feels uncomfortable driving at night.The surgeon indicated that an explant has been scheduled for (b)(6) 2019.No additional information was provided.
 
Additional Manufacturer Narrative
The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8205233
Report Number9614546-2018-01217
Device Sequence Number1746345
Product Code POE
UDI-Device Identifier05050474605701
UDI-Public(01)05050474605701(17)210809
Combination Product (Y/N)Y
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/09/2021
Device Model NumberZXT225
Device Catalogue NumberZXT225U155
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date12/28/2018
Supplement Report FDA Received Date11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-