(b)(4).Fse arrived at the site to address the reported event.Fse confirmed the errors and was able to reproduce them.Inspection of the device revealed there was a leak at the prefilter housing.Fse tightened the prefilter housing and the leak was resolved.Next, fse adjusted the flow rate to 1.10 ml/min for sa1c retention time of 0.59 minute, calibrated with calibrator lot #zs7001, and ran quality control (qc).No further issues were noted.No further action was required by field service.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 10nov2017 through aware date (b)(4) 2018.There were no similar complaints identified during the search period.The g8 operator's manual under chapter 4 screen operations & chapter 6, troubleshooting, states the following: error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.Peak pattern error: explanation: peaks were not separated well.Countermeasure: check the samples, buffers, and hemolysis & wash solution.The most probable cause of the reported event was due to a loose filter housing.
|