Model Number MMT-522LNAH
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had rf pic failed self-test multiple times.Customer¿s blood glucose was unknown.The customer was assisted with troubleshooting and reported that they were able to clear and rewind the insulin pump.The customer was advised to replace the insulin pump.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump received with motor arm failed self-test alarm during self test, problem isolated to rf board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8205656
Report Number2032227-2018-80319
Device Sequence Number1900506
Product Code OYC
UDI-Device Identifier00613994597212
UDI-Public(01)00613994597212
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522LNAH
Device Catalogue NumberMMT-522LNAH
Device Lot NumberA2522LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Initial Date Received by Manufacturer 12/20/2018
Supplement Date Received by Manufacturer03/13/2019
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date03/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
-
-