• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CLYDESDALE PTC SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CLYDESDALE PTC SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 4986655
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with spinal stenosis and radiculopathy and underwent oblique lumbar interbody fusion (olif) surgery.During surgery, the point broke off of the cage so it was no longer connected securely to the inserter.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual and optical examination identified a 3mm portion of the top face of the implant which interfaces directly with the inserter has broken off.Optical and microscopic examination of the fracture surface identified a fairly flat fracture with rays emanating from the threaded hole, witness marks on the top face of the implant and some deformation of the ribs at the top of the implant.This type of damaged is consistent with significant force during attempted implantation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLYDESDALE PTC SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key8205697
Report Number1030489-2018-01714
Device Sequence Number13928473
Product Code MAX
UDI-Device Identifier00643169190566
UDI-Public00643169190566
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K133205
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2024
Device Catalogue Number4986655
Device Lot NumberH5285339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2019
Initial Date Received by Manufacturer 12/04/2018
Supplement Date Received by Manufacturer04/19/2019
Initial Report FDA Received Date12/31/2018
Supplement Report FDA Received Date05/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-