Model Number 9529 |
Device Problem
Failure to Interrogate (1332)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the implantable cardiac monitor (icm) could not be interrogated.The icm was explanted and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The device met 80% of the lower 99% of expected longevity.Without the history of the programmed settings throughout its service life, there is no way to determine why the longevity did not match the predicted model.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The actual longevity was less than 80% of the 99.9% longevity limit.The device was fully functional, with no high current drain or evidence of battery problems.Without the history of the programmed settings throughout its service life, there is no way to determine why the longevity did not match the predicted model.If information is provided in the future, a supplemental report will be issued.
|
|