• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/16/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had critical pump error alarm.The customer's blood glucose value was 200 mg/dl.Troubleshooting was done.The customer stated that they able to saw critical pump error image on screen.Customer advised the pump will need to be replaced.Advised to discontinue use of the pump and revert to a back-up plan.The customer was advised to replace and to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device was received with a critical pump error (open book image) alarm due to moisture damage on the electronic assembly.Unable to perform the self test, displacement test, rewind test, prime test, basic occlusion test, force sensor test, occlusion test, sleep current measurement test, and active current measurement test due to critical pump error (open book image) alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key8205952
MDR Text Key131734129
Report Number2032227-2018-80551
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169656840
UDI-Public(01)00643169656840
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG1F9P5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received03/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age24 YR
Patient Weight141
-
-