| Model Number |
04.647.878 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that during an unknown procedure on (b)(6) 2018, the surgeon attempted to implant the cervical spine screw of the zero-profile variable angle (zero-p va) but it did not progress.Several attempts were made which resulted in the hexagonal part of the screw to be mistreated.The surgeon opted to use a new screw and replaced the punch to break the first cortical which was still unsuccessful.A drill with guide was used to place the screw but it was noted to have slid down with the screw.There was an x-ray image captured during the procedure which showed the screw unable to progress with the head of the screw protruding.A screwdriver was used but, according to the surgeon, it did not have enough space to continue advancing.Another attempt with a double handle screwdriver was used to advance the screw which was unsuccessful.The screw no longer moved forward or backward which is why the surgeon decided to leave it and begin closure.There was a surgical delay of 15-30 minutes reported.The procedure was completed with only two (2) inserted screws since the third screw was already mistreated.The patient status was stable.Concomitant devices: screwdriver (part: unknown, lot: unknown, quantity: 2), drill bit (part: unknown, lot: unknown, quantity: 1), zero-p va implant (part: unknown, lot: unknown, quantity: 1), double handle screwdriver (part: unknown, lot: unknown, quantity: 1).This report is for one (1) 3.7mm titanium (ti) cervical spine screw self-tapping.This is report 3 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The date received by manufacturer for the initial medwatch should be december 27, 2018.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Concomitant device/s reported: unknown screwdriver (part # unknown, lot unknown, quantity 1).
|
| |
|
Search Alerts/Recalls
|